REFRIGERATION SYSTEM造冷系统

造冷系统的设计和组装都是基于保障工艺操作不变性,整个造冷系统组装于统一个机架上,位于箱体/冷阱的后部 。选取双级半封关螺杆式压缩机,实现低温并削减了泄漏的风险,出格是在当吸气压力低于常压的低温阶段 。

每一台双级压缩机都有一个D立的循环系统,既可使用R404A造冷剂,也能够选取R507造冷剂

(本规划选取R404A) 。每个压缩机循环系统都装置有必要的节造阀门、安全阀门、过滤器、湿度批示器、仪表、油分离器、冷凝器、视镜、热气旁通阀和温度调节装置等以保障系统操作的不变性,并能保障足够的造冷能力 。造冷系统同时配有聚合发泡资料的保温 。

尽可能选取铜焊以预防泄漏,系统的泄漏率检测是以氮气打压方式进行的,如有必要使用低浓度的HFC 造冷剂 。在装置前系统内部进行彻底洗濯,并在使用前进行干燥处置 。

所有造冷循环的配置均保障在各个分歧的工艺阶段可能满足冷阱或搁板的冷量供给 。这种配置也能够保障当一套系统故障,冻干法式依然能够以较低的效能持续进行 。通过造冷剂冷却器能够确保造冷成效,在肯定前提下冷却水温可到32°C 。水冷压缩机上的热量节造阀(水量调节阀)装置在造冷剂冷凝器的入口处, 用于优化效能及削减冷却水用量 。

压缩机和造冷剂 (4)_副本.jpg

The refrigeration system is designed and assembled to ensure the specified process performance is achieved reliably. The entire refrigeration group is mounted on a frame that is positioned to the rear of the chamber/condenser. The compressors are of screw type (see item 22) of two stage, semi-hermetic compressors, providing low temperature performance and minimising the risk of leakage, especially at low temperature when the suction pressure is below atmospheric pressure.

Each compressor has a discrete circuit employing either R404 or R507HFC refrigerants (R404 used as default). Each compressor circuit is equipped with the necessary control valves, safety valves, filters, moisture indicators, gauges, oil separator, suction accumulator, condenser/liquid receiver, sight glasses, hot gas bypass valve and thermostats to enable the system to operate reliably and deliver the required refrigeration performance. Where appropriate the refrigeration circuits are insulated with polymer foam.

As far as possible brazed connections are employed and fittings such as elbows are minimised to reduce the possibility for leakage. Systems are leak checked using Nitrogen at positive pressure and low concentrations of HFC refrigerants if necessary. The systems are Nitrogen purged during brazing and dehydrated prior to operation.

To ensure the most efficient use of the installed refrigeration capacity all refrigeration circuits are configured to be able to serve either condenser or shelf cooling duties. This configuration also ensures the freeze drying process can continue, albeit at a lower The refrigerant condensers are selected to ensure refrigeration performance can be maintained with a cooling water supply up to 32°C at the specified supply rate.  Thermostatically controlled valves (water regulating valves) on the water cooled refrigerant compressors are provided at the inlet of the refrigerant condensers to optimise efficiency and minimise water usage.